A Pt100 reading of 109.73 Ω
A multimeter across a Pt100 reads 109.73 Ω. On the IEC 60751 curve that is 24.99 °C. The quick estimate, 9.73 ÷ 0.385, gives 25.3 °C, close enough to spot a faulty sensor but not to calibrate one.
Temperature 24.99 °C Open in the calculator